After cleaning, the hard question is a simple one: is it clean?
The JHPB-TLZ20 ionic contamination tester measures surface ionic concentration on printed circuit boards, electronic components, assembled products and electrical / electronic equipment. It measures in a 3:1 isopropyl alcohol / deionized water extract held at 40 ± 1 °C, shows conductivity and ionic concentration curves in real time, and outputs a report with conductivity, ionic concentration and test time when the test ends. It supports static and dynamic modes, has a measuring accuracy of ±3%, and can connect to your line data system via RS232.
Specifications
| Item | Spec |
|---|---|
| Test mode | Static / dynamic |
| Measuring range | 10 cm × 10 cm to 35 cm × 60 cm |
| Test board size | 100 mm × 100 mm to 350 mm × 600 mm |
| Accuracy (F·S) | ±3% |
| Conductivity resolution | 0.0005 (unit as given on the source page: μs/cm²) |
| Conductivity test baseline | 66–300 MΩ |
| Extract ratio | 3 (isopropyl alcohol) : 1 (deionized water) |
| Extract volume | approx. 17 L |
| Extract specific gravity | 0.85–0.855 |
| Extract working temperature | 40 ± 1 °C (constant) |
| Heater | 220 V / 1200 W heating rod |
| Power | 1300 W |
| Power supply | AC 220 V / 50 Hz (60 Hz available) / 15 A, separate earth required |
| Grounding voltage | < 0.3 V |
| Data output | RS232 (COM port); real-time conductivity, ionic concentration and time curves; report saved automatically at the end of the test |
| Dimensions | 1100 mm (L) × 600 mm (W) × 950 mm (H) |
| Weight | approx. 110 kg |
| Environment | 22 ± 2 °C, 50 ± 20% RH, normal pressure; keep away from acids, alkalis and corrosive gases or liquids |
Key advantages
Know whether it is clean only by measuring. Ionic residue on the board cannot be judged by eye after cleaning. The JHPB-TLZ20 measures conductivity and ionic concentration directly, turning cleanliness from “looks clean” into numbers you can record and trace.
Static result in 15 minutes. In static mode, a 5 ml standard solution test settles within 15 minutes; heating to 40 ± 1 °C takes 25 minutes or less; regeneration (2 ml standard solution filtered to 0.015) takes 12 minutes or less.
Large boards in one go. Boards up to 350 mm × 600 mm can be tested, so server boards and large substrates do not need several measurements.
Main components
| Item | Spec |
|---|---|
| Conductivity meter | With electrode and data sheet, 1 set |
| Ion exchange columns | 4 pcs |
| Magnetic pump | 1 unit |
| Solenoid valve | High-stability type, 1 pc |
| Heating tube | φ12 × 300, 220 V / 1200 W, 1 pc |
| Tank | PVC inner chamber 600 × 70 × 350 mm, 1 unit |
| Measuring software | Supplied with dongle |
| Computer | i3 PC with 21.5-inch monitor, 1 set |
Request a quote