中文

WTW Cond 3310 Portable Conductivity Meter | 0.000 µS/cm to 1000 mS/cm, Resistivity to 20 MΩ·cm, IP67

JIAHON is not the authorized distributor of WTW in Taiwan.

WTW Cond 3310 conductivity meter | IP67 portable, five-parameter conductivity meter

0.000 µS/cm to 1000 mS/cm / resistivity to 20 MΩ·cm / 5,000 data records / USB output

The WTW Cond 3310 is a ProfiLine series portable conductivity meter with a measuring range of 0.000 µS/cm to 1000 mS/cm. It measures five parameters — conductivity, resistivity, salinity, TDS and temperature — is IP67 waterproof, logs 5,000 data sets and transfers them by USB, covering everything from ultrapure water to high-salinity wastewater with one instrument.

Main specifications

ItemDetails
ParametersConductivity, resistivity, salinity (IOT table), TDS, temperature
Conductivity range0.000 µS/cm to 1000 mS/cm (auto-ranging)
ResistivityUp to 20 MΩ·cm (ultrapure water measurement)
Temperature range−5.0 to +105.0°C
Accuracy (conductivity)±0.5% (depending on compensation mode)
Accuracy (temperature)±0.1°C (NTC 30) / ±0.3°C (PT 1000)
Cell constant range0.010 to 25.000 cm⁻¹ (fixed, calibrated, adjustable)
Data loggingManual 200 sets / automatic 5,000 sets (with date, time, ID)
InterfaceWaterproof USB 1.1, .csv output
ProtectionIP67 (fully waterproof)
Power / battery life4 × AA batteries; 800 h without backlight / 100 h with backlight
Dimensions / weight180 × 80 × 55 mm / approx. 0.4 kg

Applications

ApplicationWhy it fits
Semiconductor / ultrapure water monitoringResistivity measurement up to 20 MΩ·cm, accuracy ±0.5%
Field water quality inspectionIP67 waterproof + light 0.4 kg, easy to carry, splash-proof
Process water records5,000 automatic records + USB to PC for water quality reports
Boiler / cooling water monitoringTemperature measurement to 105°C, conductivity range up to 1000 mS/cm
Food / beverage plants, check after CIPAutoRead records only when stable, for reliable cleaning verification data
Environmental water sample analysisSalinity (IOT table) + TDS, multiple parameters in one instrument

Operating features

  • AutoRead: automatically detects a stable reading, preventing premature recording while the value is still drifting
  • HOLD: freezes the display after measurement for easy recording and screenshots
  • Temperature compensation: none, linear (lin) and nonlinear (nLF, per EN 27888), with reference temperature selectable at 20°C or 25°C
  • Backlit graphic LCD: clearly readable in bright outdoor light, for field use

Email us for a quote: [email protected]