A temperature humidity test (also called damp heat testing) places samples in an environmental test chamber — a temperature/humidity chamber, called 環境試驗機 or 恆溫恆濕試驗箱 in Taiwan — under steady or cyclic temperature and relative humidity, then checks electrical, visual and functional degradation. “85/85” means 85°C and 85%RH, the condition of the JEDEC JESD22-A101 temperature-humidity-bias (THB) life test, typically run for 1,000 hours (sources 1, 2).
Before anything else, write the test as a table someone else could rerun: temperature, humidity, ramp rate, dwell, cycles or hours, bias, and checkpoints. If one field is missing, results cannot be compared.
1. Common standards and their conditions
| Standard (current revision) | Test | Conditions | Source |
|---|---|---|---|
| JESD22-A101 (D.01) | THB | 85°C / 85%RH, 1,000 h, biased | 1, 2 |
| JESD22-A110 (E.01) | HAST | 130°C / 85%RH for 96 h, or 110°C / 85%RH for 264 h, biased (values from rev. B) | 3, 2 |
| JESD22-A102 | Autoclave / PCT | 121°C / 100%RH, 168 h, unbiased | 2 |
| JESD22-A104 | Temperature cycling | Transfer ≤ 1 min, dwell ≥ 10 min, reach temperature within 15 min; e.g. condition B −55 to 125°C | 2 |
| JESD22-A106 (B.02, Jan 2023) | Thermal shock, liquid-to-liquid | Transfer ≤ 10 s, dwell ≥ 2 min, reach temperature within 5 min; condition C −55 to 125°C, 15 cycles (values from rev. A) | 4, 5, 2 |
| IEC 60068-2-78:2025 (ed. 3) | Test Cab, damp heat steady state | Constant temperature, high humidity, no condensation on the specimen; check severity values in the ed. 3 text | 6 |
| IEC 60068-2-30:2025 (ed. 4) | Test Db, damp heat cyclic (12 h + 12 h) | Ed. 3 values: upper temperature 40°C (2–56 cycles) or 55°C (1, 2, 6 cycles), 93 ± 3%RH at the upper temperature, ≥ 95%RH during the rise, lower temperature 25 ± 3°C; condensation may occur during the rise | 7, 8 |
What is JESD22-A106? JEDEC’s thermal shock method. Parts move between two liquid baths (water or perfluorocarbon; perfluorocarbon is required for conditions B, C and D). If interruptions exceed 10% of the specified cycles, the test restarts (source 5). Air-to-air two-zone shock is usually referenced to JESD22-A104 or IEC 60068-2-14 Test Na; the transfer and soak rules differ, so the methods are not interchangeable.
JEDEC conditions target packaged components and are severe. For boards, modules and finished products, start from the customer’s specification and the real use environment, and record why a given test was chosen.
2. Dew point and condensation — the calculation
Condensation forms when a surface is at or below the air’s dew point. The Magnus approximation:
γ = ln(RH/100) + a·T/(b + T), Td = b·γ/(a − γ), with a = 17.62 and b = 243.12°C.
- 85°C / 85%RH: γ = −0.163 + 4.565 = 4.402, so Td ≈ 81°C.
- 23°C / 50%RH room air: Td ≈ 12°C.
Hypothetical case: the chamber air has reached 85/85, but a thermocouple on a heavy-copper area reads 60°C. Since 60 < 81, water condenses on the board. That is why steady damp heat should stabilize temperature before raising humidity, and why samples pulled from a cold test must warm up before measurement in a 23°C/50%RH room.
3. Seven fields every test condition needs
- Temperature and humidity (high, low, and the temperature at which RH is specified; starting point, often 25°C/50%RH)
- Ramp rate, and whether it refers to chamber air or the sample
- Dwell, counted from when the sample reaches temperature
- Cycles or hours, with interim readouts
- Bias (voltage, load, schedule; JESD22-A110 gives continuous vs. cycled bias rules, source 3)
- Loading (quantity, shelf position, wall clearance — do not overfill)
- Checkpoints and pass criteria, written before the test
4. Reading an environmental chamber datasheet
- Humidity range is an area, not a line. It is bounded by dew point and dry-bulb temperature. Confirm that your setpoint sits inside the chamber’s temperature/humidity envelope.
- Ramp rates are usually measured empty, without humidity control. Loaded and powered samples slow them down.
- Fluctuation is not uniformity. IEC 60068-3-5 (temperature chambers) and IEC 60068-3-6 (temperature/humidity chambers), both edition 2 (2018), confirm chamber performance without specimens and treat fluctuation, gradient and variation in space separately (sources 9, 10). A study applying the IEC procedure defines fluctuation as the change from the mean at one point after stabilization, and gradient as the largest difference in mean value between two points at any moment (source 11). A “control accuracy ±0.5°C” usually describes fluctuation at the control sensor; “uniformity ±1.0°C” describes spatial difference.
- Heat load and volume. Add up the dissipation of powered samples and compare with the chamber’s load capacity. One test lab notes that for heat-dissipating equipment in steady damp heat, chamber volume should be at least five times the specimen volume (source 12).
Jiahon’s JHHWS-260BE temperature humidity chamber lists 260 L, 30–98%RH and 24-segment programming. The JHHWS-260BE covers 0–70°C; for special conditions such as 85°C/85%RH, please contact us.
5. Example test plan (hypothetical)
- Product: control module for an outdoor cabinet, reflowed, not potted.
- Goal: compare insulation degradation of two conformal coatings.
- Condition: 85°C/85%RH for 1,000 h, one unbiased and one 12 V biased group; readouts at 0, 168, 500 and 1,000 h.
- Samples: 3 lots × 5 boards per coating, plus 3 uncoated controls.
- Loading: up to 6 boards per shelf, boards parallel to airflow, ≥ 10 cm from walls.
- Measurement: one Type K thermocouple on a heavy-copper area per board, one in chamber air, plus an independent data logger.
- Pass criteria (hypothetical): all functional tests pass; surface insulation resistance not below the specified value; no corrosion or blistering at 10×.
- Recovery: at each readout, let boards return to room temperature with dry surfaces before measuring, using a fixed recovery time.
6. Where this applies
SMT lines and assembly test (material changes such as coatings, solder paste or flux), incoming inspection of connectors and cables, ESD control (surface resistance of ESD packaging at high or low humidity), and cleanroom consumables (appearance and particle changes).
7. Common mistakes
- Trusting the chamber display instead of a thermocouple on the sample body.
- Attaching sensors to cables or fixtures rather than to slow- and fast-heating spots on the sample.
- Raising humidity and temperature together in a steady damp heat test.
- Treating condensation in IEC 60068-2-30 cyclic tests as a fault — it may occur by design.
- Ignoring sensor uncertainty: Type K class 1 tolerance is ±1.5°C or ±0.4%, whichever is greater (IEC 60584-1), larger than many chambers’ control spec.
- No rule for interruptions written in advance.
- Opening the door frequently, which causes fluctuation, condensation or even icing.
8. FAQ
What does 85% mean in high-humidity testing? 85% relative humidity, usually paired with 85°C as “85/85”, the JESD22-A101 THB condition (sources 2, 13). JESD22-A110 states that parts reaching moisture equilibrium within 24 h under HAST are equivalent to at least 1,000 h at 85/85 (source 3).
Which standards cover temperature humidity testing? JESD22-A101 and JESD22-A110 for packaged devices (sources 1, 3); IEC 60068-2-78 and IEC 60068-2-30 for components and equipment (sources 6, 7).
What is JESD22-A106? JEDEC’s liquid-to-liquid thermal shock method, current revision B.02 (January 2023) (source 4).
Sources are listed in the Chinese section. Standards must be purchased from their publishers; values marked as from earlier revisions should be checked against the current edition. This article is general information and does not endorse any product or process; follow your customer’s specification and the component manufacturer’s datasheet.